A Calibration Method Based on Dual Quadratic Fitting and Nine Parameters for Mining Non-Cooled Infrared Thermal Imager

Infrared detection method is mainly used to detect infrared energy field, through which the spontaneous combustion zone of coal can be comprehensively judged. In order to ensure the accuracy of the temperature of the test target under the mine, the paper mainly studies the calibration method of the uncooled infrared imager under the influence of the environment temperature, distance, the effect of the working temperature of the detector and the other factors. A new method of calibration model of dual quadratic fitting and the calibration equation of nine parameters method are proposed. Compared to existing methods, the proposed method is used to process the experiment, and it is concluded that the average error is within 1°C, which can meet the requirements of temperature measurement-thermal imaging under the mine.