Study of Influence of Hole Distance on the Strength of Locking Compression Plate Implants

Locking Compression Plate (LCP) implants have become an increasingly popular choice for orthopaedic surgeries, particularly in treating tibia bone fractures. Tibia implants have a failure rate from 7% to 18%. This study aimed to investigate the influence of the distance between the holes in the LCP implants on their overall strength and mechanical stability. The researchers employed a finite element analysis approach to simulate the performance of LCP implants made of Ti6Al4V Titanium Alloy. As Ti6A14V Titanium alloy has superior corrosion resistance, biocompatibility, and lower modulus of elasticity, which distribute stress evenly, we have selected this material for our study. This comparative analysis suggests that increasing the distance between the holes from 5mm to 8mm in the LCP implant can enhance mechanical stability and load-bearing capacity. Finite element analysis suggested that the increase in hole distance from 5mm to 8mm reduced the stress value from 89.541 MPa to 81.734 MPa. The stresses in the implant increased to 88.025 MPa after increasing the hole distance beyond 8mm. These insights can inform the optimization of LCP implant design, particularly in treating tibia bone fractures, to ensure secure fixation and improved patient clinical outcomes. Major Findings: Increasing the hole distance in the LCP implant from 5 mm to 8 mm enhances mechanical stability and reduces stress from 89.541 MPa to 81.734 MPa. However, increasing the distance beyond 8 mm raises the stress to 88.025 MPa, making the implant unstable. The inverse correlation exists between the hole distance and maximum stress magnitude.