DPP Determination of Cadmium after Adsorption of its Piperazine Dithiocarbamate Complex onto Microcrystalline Naphthalene

Cadimium piperazine dithiocarbamate complex is quantitatively adsorbed onto micrystalline naphthalene in the pH range of 3.4 - 11.0. Cadmium-piperazine dithiocarbamate-naphthalane complex is filtered, shaken with 10 ml of 1 M HCl and cadmium determined polarographically. The condition for its estimation in complex materials is optimised. Linearity exists in the concentration range of 0.07-10 ppm with a correlation factor of 0.9997. Ten replicate determinations of a sample solution containing 112 pg of cadmium gave a relative standard deviation of +0.78%. The dctection limit is 0.07 ppm (signal to noise ratio = 2).