Electrochemical Evaluation Of Hydroxyapatite/Halloysite Nanotube/Polyvinyl Pyrrolidone Composite Coating On TI6AL-4V Alloy For Orthopedic Applications

Excellent mechanical properties, corrosion resistance and favorable biocompatibility of Ti-6Al-4V alloys have made them to receive persistent attention as orthopedic implant. However, its inadequate osseointegration property is a major hindrance for such applications and for this reason, it is important to introduce bioceramicpolymer coatings on the metallic surface to attain true adhesion between the implant and bone. Thus, the present study deals with the development of composite coatings of HAP/HNT/PVP on Ti-6Al- 4V alloy by electrophoretic deposition method. The coatings were characterized Fourier Transform Infrared Spectroscopy (FTIR) which revealed the presence of functional groups. Scanning Electron Microscopy (SEM) demonstrated the morphology. The elemental analysis were done by Electron Dispersive X-Ray Analysis (EDAX). The results of potentiodynamic studies in the simulated body fluid solutions showed that the HAP/HNT/PVP coated Ti-6Al-4V provided better corrosion protection. Hence, HAP/HNT/PVP composite coated Ti-6Al4V alloy might be a promising material for fabrication of bone implant

View article page

Download PDF