Synergistic electrocatalytic Action Of Ploy (P-Toluene Sulphonic Acid) And Gold Nanoparticles Based Glassy Carbon Electrode for The Simultaneous Voltammetric Determination Of Xanthine And Hypoxanthine

Xanthine (XA) and hypoxanthine (HX) are the purine bases which are the important intermediates in the purine metabolic pathway. They are mainly found in body tissues and fluids. Hence their abnormal levels are indicative of many clinical conditions such as xanthinuria, hyperuricemia, gout and renal failure. Thus their determination is important from the clinical diagnostic point of view. Electrochemical sensors offer rapid, reliable, less expensive, sensitive and selective simultaneous determination. In view of this, attempt has been made to develop a poly (para toluene sulfonic acid)/gold nanoparticles modified glassy carbon electrode (pPTSA /AuNPs /GCE) as a sensor for the simultaneous determination of XA and HX. The experimental parameters for the simultaneous determination of XA and HX on pPTSA /AuNPs /GCE has been optimised. The lowest potential for the electro-oxidation of XA and HX was obtained in 0.1 M NaOH. The increase in peak current with concentration was linear for individual as well as simultaneous determination of these analytes. For simultaneous determination, peak current changes with concentration in the range 6-700 µM for XA and 40-600 µM for HX. The mechanistic and kinetic parameters were determined. The sensor was successfully applied for determination of XA and HX in artificial urine and serum.