A Comparative Evaluation of Electrostatic Discharge Testing Requirements in IEC 61000-4-2:2025 and ISO 10605:2023
Electrostatic Discharge (ESD) remains one of the causes of failures in modern electronics, especially in microelectronics. This paper provides a critical comparative analysis of the recently released IEC 61000- 4-2:2025 and the updated ISO 10605:2023. Both standards define immunity requirements. The recent revisions introduced methodologies for calibration, waveform verification, and application-specific testing, such as for wearable products and Electric Vehicles (EVs). This study analyses the various requirements of both standards with respect to test levels, test equipment requirements, waveform specifications, calibration requirements of the ESD generator (with Air and Contact Discharge Tips), verification aspects of the ESD test setup, test setup and test procedure, Measurement Uncertainty (MU) considerations, test setups specified for wall-mounted equipment and wearable devices, compliance criteria, etc. Furthermore, it evaluates and highlights the significant technical changes in IEC 61000-4-2:2025, specifically the introduction of second peak current (Ip2), the standardized calibration for air discharge tips, and the introduction of Annex J for wearable devices, along with other technical changes in the standard. The novelty of this study is to highlight the provisions of both the latest published standards by comparing them.