Electrophoretically Modified 316L Stainless Steel: A Promising Approach for Durable and Biocompatible Bone Implants
Implant failures often stem from poor osseointegration, electrochemical manifestations, and infections. This research addressed these challenges by developing a bioceramic-based composite coating for Stainless Steel (316L SS), aiming to prevent the release of toxic metal ions in the body. Nano-hydroxyapatite (nHAP) was synthesised using a wet chemical precipitation method, while Graphene Oxide (GO) was prepared via a modified Hummers' method. A 99:1 nHAP-GO composite was then created using an ex-situ method. Both nHAP and the 99:1 nHAP-GO composite were coated onto 316L SS substrates through Electrophoretic Deposition (EPD), utilising isopropanol as the solvent. Characterisation techniques such as Fourier transform infrared (FTIR) and X-ray Diffraction (XRD) confirmed the presence of functional groups, phase purity, and the crystalline nature of the synthesised and prepared samples. Atomic Force Microscopy (AFM) analysis revealed that the coated samples had increased roughness compared to bare 316LSS. This enhanced roughness is beneficial, as it promotes a better osseointegration process for the coated implants. Crucially, electrochemical studies demonstrated significantly higher corrosion resistance for the composite-coated 316LSS when compared to both nHAPcoated 316LSS and bare 316LSS.